Vista de página 0
2216 60th Drive East • Bradenton, Florida 34203
(941) 751-7596 • Fax (941) 751-7598 • www.plastecvent.net
PLASTEC
®
VENTILATION, INC.
Specifications and Owner’s Manual
Laboratory Airflow Monitors & Controls
TYPE A STANDARD &
NIGHT SETBACK
LED & DIGITAL
Vista de página 0
1 2 3 4 5 6 ... 13 14

Resumo do Conteúdo

Página 1 - LED & DIGITAL

2216 60th Drive East • Bradenton, Florida 34203(941) 751-7596 • Fax (941) 751-7598 • www.plastecvent.netPLASTEC® VENTILATION, INC.Specifications

Página 2 - General Description

Front plate/fasciaChemical and pressure resistant stickerFlashing red LED for alarm: air speed too lowAlarm Mute push buttonResin StickerDigital displ

Página 3 - Operating Principle

Front plate/fasciaChemical and pressure resistant stickerFlashing red LED for alarm: air speed too lowAlarm Mute on/offpush buttonCustomizableResin St

Página 4

INSTALLATION1) Drill a Ø 23mm / 0.91" hole either in the top or side of the fume cupboard to allow sensor positioning. Make sure sensor is not in

Página 5

VELOCITY TYPETo change Digital readout from/to m/s to FPM:Press simultaneously the following buttons: until buzzer sounds (about 5 seconds).DISP

Página 6

MAINTENANCEFront plate and stickers of airflow monitor may be cleaned with mild soap and water on a damp cloth to remove finger marks, oils and residu

Página 7

General Description2Laboratory Airflow Monitors & ControlsTYPE A STANDARD & NIGHT SETBACKLED & DIGITALBENEFITS• Safety: in accordance wit

Página 8 - Contacts & Features

When the fume extraction fan is running, it causes negative pressure inside the fume cupboard. If the sash is lowered, the negative pressure becomes m

Página 9 - Specifications

115 VAC /12 VDCPower SupplyControllerSensorSensorSignal InputSash High Alarm InputSash movement=Pressure decrease or increaseHood Lighton/off relayAir

Página 10 - TYPE A STANDARD

115 VAC /12 VDCPower SupplyControllerSensorSensorSignal InputSash High Alarm InputSash movement=Pressure decrease or increaseHood Lighton/off contactN

Página 11 - TYPE A NIGHT SETBACK

5SpecificationsAirControl A LEDPart Number 819700 819701 1 green LED for normal air speed 1 green LED for normal air speed 1 red LED flashing for

Página 12

AirControl A LEDPart Number 819700 819701 1 green LED for normal air speed 2 green LED for normal air speedDisplay – Visual 1 red LED flashing fo

Página 13

Alarm Relay220v 3A(optional)ONOFFSettings:1. On: Alarm always active Off: Alarm disabled when fan is off2. Alarm delay setting: On: 30 seconds Off:

Página 14

MONITOR & MOUNTING BOX35 mm / 1.38"24 mm / 0.94"Mounting Box90 mm / 3.54"210 mm / 8.27"SENSOR65 mm / 2.56"28 mm / 1.10&qu